Reseach Article

Comparison of 2-D Algorithms in ElT based Image Reconstruction

by Vidya Sarode, Sneha Patkar, Alice N Cheeran
International Journal of Computer Applications
Foundation of Computer Science (FCS), NY, USA
Volume 69 - Number 8
Year of Publication: 2013
Authors: Vidya Sarode, Sneha Patkar, Alice N Cheeran
10.5120/11860-7642

Vidya Sarode, Sneha Patkar, Alice N Cheeran . Comparison of 2-D Algorithms in ElT based Image Reconstruction. International Journal of Computer Applications. 69, 8 ( May 2013), 6-11. DOI=10.5120/11860-7642

@article{ 10.5120/11860-7642,
author = { Vidya Sarode, Sneha Patkar, Alice N Cheeran },
title = { Comparison of 2-D Algorithms in ElT based Image Reconstruction },
journal = { International Journal of Computer Applications },
issue_date = { May 2013 },
volume = { 69 },
number = { 8 },
month = { May },
year = { 2013 },
issn = { 0975-8887 },
pages = { 6-11 },
numpages = {9},
url = { https://ijcaonline.org/archives/volume69/number8/11860-7642/ },
doi = { 10.5120/11860-7642 },
publisher = {Foundation of Computer Science (FCS), NY, USA},
address = {New York, USA}
}
%0 Journal Article
%1 2024-02-06T21:29:39.486851+05:30
%A Vidya Sarode
%A Sneha Patkar
%A Alice N Cheeran
%T Comparison of 2-D Algorithms in ElT based Image Reconstruction
%J International Journal of Computer Applications
%@ 0975-8887
%V 69
%N 8
%P 6-11
%D 2013
%I Foundation of Computer Science (FCS), NY, USA
Abstract

In Electrical Impedance Tomography (EIT), the resulting voltages developed by injecting a constant electrical current at the boundary electrodes are used to obtain images representing conductivity. Electrical Impedance and Diffused Optical Reconstruction Software (EIDORS) is used in this tomographic imaging. This paper compares different reconstruction algorithms used in EIT experiments conducted on a circular plastic phantom. Priors like Noser, Laplace, Tikhonov, Total Variation (TV) and Gauss-Newton and Back Projection algorithms are used for detection of nonconducting impurity. The images thus obtained are compared with respect to shape and size of impurity and perseverance of image contrasts.

References
  1. Tushar K. Bera, J Nagaraju, "Electrical Impedance Tomography (EIT) System for Radiation-Free Medical Imaging Based on LabVIEW", pp. : 1-4.
  2. Bera T. K and Nagaraju J. , "Resistivity Imaging of a reconfigurable phantom with circular inhomogeneities in 2D-electrical impedance tomography", Measurement 44(2011), pp 518-526, doi:10. 1016/j. measurement . 2010. 11. 015.
  3. Erwati M. T, and Farrukh N. , "Applications of electrical impedance tomography for imaging in biomedical and material technology", Proceedings of 2009 IEEE Students Conference on research and development, Nov '09.
  4. Andy Adler and William R. B. Lionheart, "Uses and abuses of EIDORS: An extensible software base for EIT", The University of Manchester, May 2006.
  5. Amir Movafeghi, Ali R. Nateghi, ManouchehrSoleimani, Mohammad H. Kargarnovin and Hamid Soltanianzadeh, "Image Reconstruction Algorithms or Sut-1 Eit System", pp. : 579-582, 2004.
  6. Nicholas Polydorides, "Image Reconstruction Algorithms for Soft-field tomography", Manchester, United Kingdom, September 2002.
  7. VidyaSarode, Priya M. Chimurkar, Alice N Cheeran, "Electrical Impedance Tomography using EIDORS in a Closed Phantom", International Journal of Computer Applications (0975 – 888), Volume 48– No. 19, June 2012.
  8. TusharKantiBera and J. Nagaraju, "A Study of Practical Biological Phantoms with Simple Instrumentation for Electrical Impedance Tomography (EIT)", I2MTC, Singapore, 5-7 May 2009.
  9. Bera T. K and Nagaraju J. , "Studying the resistivity imaging of chicken tissue phantoms with different current patterns in Electrical Impedance Tomography (EIT)", Measurement (2012), doi:10. 1016/j. measurement . 2012. 01. 002.
  10. Zou Y. and Guo Z. , "A review of electrical impedance techniques for breast cancer detection", Elsevier Medical Engineering and Physics 25, (2003).
  11. D. C. Barber, N. J. Avis, "Single Step Algorithms for Image Reconstruction", Electrical Impedance Tomography Applied Potential Tomography IEE Colloquium on (1992), pp 4/1-4/3.
  12. TusharKantiBera, J. Nagaraju, "A Chicken Tissue Phantom for Studying an Electrical Impedance Tomography (EIT) System Suitable for Clinical Imaging", Sens Imaging (2011) 12:95–116 DOI 10. 1007/s11220-011-0063-4.
  13. Andrea Borsic, William R. B. Lionheart, and Christopher N. McLeod, "Generation of Anisotropic-Smoothness Regularization Filters for EIT", IEEE Transactions on Medical Imaging, Vol. 21, No. 6, June 2002.
  14. M. Cheney and D. Isaacson, "NOSER: An Algorithm for Solving The Inverse Conductivity Problem", International Journal of Imaging Systems and Technology, Vol. 2, 66-75 (1990).
  15. Borsic, B. M. Graham, A. Adler , W. R. B. Lionheart, "In vivo Impedance Imaging with Total Variation Regularization", pp. : 1-10.
Index Terms

Computer Science
Information Sciences

Keywords

Priors Regularization Impurities Jacobian